Metrology and calibration

Clear explanations of concepts, methods, standards and measurements for engineers and technicians.

  • What does “bias” mean in metrology?

    Bias in metrology is the difference between the average of many repeated indications and the reference value; it is the systematic part of the error, which is not reduced by averaging more readings.

  • What does “measurand” mean?

    The measurand is the particular quantity to be measured, defined with sufficient precision: not “the temperature”, but the temperature of a certain fluid, at a specific point and under specific conditions.

  • What Does ISO 10012 Set Out for Measurement Processes?

    This standard sets out the requirements for a measurement management system: how an organization controls its measurement processes and equipment to ensure they are suitable for the intended use. Consult the current edition.

  • What does the Federal Law on Metrology and Standardization establish?

    The Federal Law on Metrology and Standardization (Ley Federal sobre Metrología y Normalización) is the Mexican legal instrument that regulates the system of units, metrology, standardization, conformity assessment and accreditation. For its specific provisions, consult the current text.

  • What is a decision rule with a guard band?

    A decision rule with a guard band accepts or rejects a measurement result using an acceptance limit offset from the tolerance limit, so that measurement uncertainty reduces the risk of accepting an out-of-tolerance instrument.

  • What Is a Multifunction Calibrator?

    A multifunction calibrator is an instrument that generates electrical and process signals with good accuracy, such as voltage, current, resistance and sensor simulation, and sometimes measures these quantities, in order to calibrate and test other instruments.

  • What Is a Reference Material?

    A reference material is a material sufficiently homogeneous and stable with respect to one or more properties, with assigned values and their uncertainty, used to calibrate, validate methods or control the quality of measurements.

  • What Is a Reference Multimeter Used For?

    A reference multimeter serves as a working standard for measuring voltage, current and resistance with low uncertainty, so that electrical equipment and sources of lower accuracy can be verified or calibrated.

  • What Is a Standard Resistor?

    A standard resistor is a resistor built to have a known, very stable resistance value with a low temperature coefficient, used as a reference in the calibration of resistance meters and other instruments.

  • What is a tolerance?

    A tolerance is the total permitted variation of a quantity around its nominal value; it defines how far a part or process can be from the desired value and still be considered acceptable.

  • What Is a Voltage Reference Source?

    A voltage reference source is a device that delivers a very stable, known voltage, used as a standard or as a point of comparison to calibrate voltmeters, converters and other instruments.

  • What is a working standard?

    A working standard is the standard used routinely to calibrate or verify instruments, and which is in turn calibrated against a higher-level standard to maintain traceability.

  • What is an uncertainty budget and how is it built?

    An uncertainty budget is the ordered list of all the sources contributing to the doubt about a measurement result, with their magnitude and weight, which are combined according to a mathematical model of the measurement to obtain the combined and expanded uncertainty.

  • What is measurement uncertainty?

    Measurement uncertainty is a parameter that characterizes the dispersion of the values that can reasonably be attributed to the measured quantity. It is not an error: it is the quantified doubt that accompanies a result.

  • What is metrological traceability?

    Metrological traceability is the property of a measurement result whereby it can be related to a reference, normally a national or international standard, through an unbroken, documented chain of calibrations, each with its own uncertainty.

  • What is normalized error in an interlaboratory comparison?

    Normalized error compares the difference between a laboratory's result and the reference value with the combined uncertainty of both; a value whose magnitude does not exceed one indicates agreement within what was stated.

  • What Is Reproducibility in a Measurement?

    Reproducibility is the precision of a measurement when conditions change: a different operator, instrument, laboratory, location or time, even though the same thing is measured with the same method.

  • What is the coverage factor?

    The coverage factor is the number by which the combined standard uncertainty is multiplied to obtain the expanded uncertainty, that is, an interval with a stated level of confidence.

  • What is the difference between accuracy, precision and repeatability?

    Accuracy indicates how close a measurement is to the true value; precision, how close several repeated measurements are to one another; repeatability is precision under the same conditions and over a short period.

  • What is the GUM uncertainty guide?

    The GUM is an international document that sets out general rules for evaluating and expressing measurement uncertainty, with a method for combining contributions and stating a result with its expanded uncertainty.

  • What Is the Hierarchy of Measurement Standards?

    The hierarchy of measurement standards is the ordering of standards by level of accuracy, from the realization of the unit at a national institute down to working instruments, so that each level is calibrated against the one immediately above it.

  • What Is the International System of Units?

    The International System of Units, or SI, is the measurement system used worldwide. It is based on seven base units, defined from constants of nature, from which all other units are derived.

  • What is the International Vocabulary of Metrology?

    The International Vocabulary of Metrology, known as the VIM, is the document that defines the fundamental terms of metrology, such as quantity, calibration, traceability and uncertainty, so that they are used with the same meaning everywhere.

  • What is the maximum permissible error of an instrument?

    The maximum permissible error is the extreme value of the indication error that a specification, standard or requirement accepts for an instrument. If the measured error, taking its uncertainty into account, exceeds that limit, the instrument does not comply.

  • When is a rectangular distribution used in uncertainty estimation?

    A rectangular distribution is used when only the limits within which the value lies are known and there is no reason to think some values are more likely than others; the standard uncertainty is then the half-width divided by the square root of three.

  • Why do two calibrations of the same instrument not match?

    Two calibrations of the same instrument can differ because of real instrument drift, because of differences in method, points, conditions and standards, or simply because the difference fits within the stated uncertainties.

  • Why does a calibrated instrument still give wrong readings?

    A calibrated instrument can give wrong readings because calibration only describes its state at one moment and under specific conditions: drift, damage, misuse, environmental effects, incorrect installation or an unsuitable method can all appear afterward.

  • Why does a calibration certificate lose its validity?

    A calibration certificate has no expiration date of its own; it loses its practical validity when the instrument changes state, when the interval set by the user or by the applicable requirement runs out, or when the standard used is no longer traceable.

  • Why does a measurement standard deteriorate?

    A standard deteriorates through wear, contamination, corrosion, component aging, impacts, improper handling, and storage or use conditions outside those intended, all of which shift its value from the one on the certificate.