Metrology and calibration

Clear explanations of concepts, methods, standards and measurements for engineers and technicians.

  • Calibration or Adjustment: Are They the Same?

    They are not the same: calibration compares the instrument with a standard and documents its error; adjustment modifies the instrument to reduce that error. You can calibrate without adjusting, and adjust without having validly calibrated.

  • Calibration or Verification: Which Do You Need?

    Calibration is needed when you must know the instrument's error with its uncertainty and establish traceability; verification is enough when you only need to confirm that it meets a requirement, without rigorously quantifying the error.

  • Comparison Calibration or Direct Calibration?

    In direct calibration, the instrument measures or generates a value that is immediately compared with a standard; in comparison calibration, both receive the same quantity at the same time and their indications are compared, with no need to apply an exact known value.

  • How Are Environmental Effects Compensated for in a Calibration?

    Environmental effects are compensated for by controlling the conditions, applying corrections calculated from known models or coefficients or, if that is not possible, including their contribution in the calibration uncertainty.

  • How are out-of-tolerance instruments managed?

    An out-of-tolerance instrument is managed by removing it from use, assessing which measurements it may have affected since its last good verification, deciding whether to adjust, repair or retire it, and reviewing why it deviated.

  • How Are Results Compared Between Laboratories?

    Results between laboratories are compared by measuring the same item or material with the same protocol and evaluating whether the differences are compatible with the stated uncertainties, normally through proficiency testing or interlaboratory comparisons.

  • How Do You Choose a Standard to Calibrate an Instrument?

    A standard is chosen so that its uncertainty is clearly smaller than that of the instrument being calibrated, it covers the required range and quantity, it is compatible with the method and it has a current traceable calibration.

  • How Do You Decide Whether an Instrument Conforms Using Uncertainty?

    To decide conformity, the measured result is compared with the tolerance limit by applying an agreed decision rule that takes measurement uncertainty into account and defines when the result is accepted, rejected or left undetermined.

  • How does instrument resolution enter into the uncertainty?

    Instrument resolution enters the uncertainty as a Type B component with a rectangular distribution: its half-width is half the resolution, and the standard uncertainty is that half divided by the square root of three.

  • How is a control chart used to monitor a working standard?

    A control chart for a working standard periodically records the measurement of a stable check standard and compares each point with limits calculated from its own history, to detect drift, jumps or increases in spread between calibrations.

  • How is a digital multimeter calibrated in general terms?

    A digital multimeter is calibrated by applying known signals, generated by a traceable calibrator or standard, on each selected function and range, and comparing the multimeter reading with the reference value to obtain the error and its uncertainty.

  • How is a measuring equipment control program organized?

    A measuring equipment control program is organized around what the process needs to measure, not around the instruments available: take inventory, define the requirement for each measurement, confirm that the equipment meets it and maintain that evidence over time.

  • How Is a Repeatability and Reproducibility Study Done?

    A repeatability and reproducibility (R&R) study measures representative parts several times with several operators and separates the variation of the measurement system from the real variation between parts, to show whether the system is adequate.

  • How Is an Adjustment Performed After a Calibration?

    An adjustment after calibration is made by modifying the instrument so that its indication moves closer to the reference value. It must always be preceded by recording the error found and followed by a verification of the result.

  • How is an instrument prepared for calibration?

    Preparing an instrument for calibration means delivering it clean, complete, stable and without prior adjustment, along with information on its use, so that the calibration reflects how it really behaves and not how it was left after being handled.

  • How Is an Instrument’s Calibration Interval Determined?

    The calibration interval is not set by a universal rule: it is determined from the stability of the instrument, the risk of an incorrect measurement, its use and its history of results, and it is adjusted over time.

  • How Is Linearity Determined?

    Linearity is determined by measuring the instrument's response at several points across the span, fitting a reference line to the data and calculating the maximum deviation of the points from that line.

  • How Is Long-Term Drift Determined?

    Long-term drift is determined by comparing the results of successive calibrations of the same instrument over time and analyzing how the error at each point changes from one calibration to the next.

  • How is metrology managed on a production line?

    On a production line, metrology is managed by identifying which product characteristics depend on a measurement, ensuring that each measurement is capable enough for its tolerance, and continuously verifying that it remains so.

  • How Is the Hysteresis of an Instrument Measured?

    Hysteresis is measured by applying the same values of the quantity in increasing and then decreasing order and comparing the indications at each point: the difference between the two paths is the hysteresis.

  • How Is the Indication Error of an Instrument Determined?

    Indication error is determined by comparing the instrument reading with the value of a traceable standard at several points across the span; the error is the indication minus the reference value, and it is reported together with its uncertainty.

  • How Is the Stability of a Reference Standard Verified?

    It is verified by comparing the results of successive calibrations and verifications of the standard over time, to estimate how much it drifts and whether that drift fits within what the uncertainty of the work allows.

  • How Is the Uncertainty of a Calibration Estimated?

    The uncertainty of a calibration is estimated by identifying the sources of variation, expressing each one as a standard uncertainty, combining them according to a measurement model and multiplying by a coverage factor to obtain the expanded uncertainty.

  • How is the uncertainty of an indirect measurement evaluated?

    In an indirect measurement, uncertainty is evaluated by combining the uncertainties of the input quantities weighted by their sensitivity coefficients, which are the partial derivatives of the function that defines the measurand.

  • How is the validity of results ensured between calibrations?

    The validity of results between calibrations is ensured with intermediate checks, check standards, drift tracking, comparisons with other equipment or laboratories, and good conditions of use and storage.

  • Manufacturer’s Tolerance or In-House Acceptance Criteria?

    The manufacturer's tolerance describes what the instrument can do under specified conditions; an in-house acceptance criterion is defined by what your process needs to measure. You should apply the more demanding one that corresponds to your use.

  • What Are a Primary Standard and a Secondary Standard?

    A primary standard is one established with a method of the highest metrological quality without depending on another standard of the same quantity; a secondary standard receives its value by comparison with a primary one.

  • What Are Systematic and Random Error?

    Systematic error is the component of error that remains constant or varies predictably when a measurement is repeated, and it can be corrected if known; random error varies unpredictably and is reduced by averaging.

  • What causes an instrument to go out of tolerance between calibrations?

    An instrument goes out of tolerance between calibrations because of component drift, damage from impacts or overloads, adverse ambient and operating conditions, contamination, wear or improper intervention. Periodic calibration is what detects it.

  • What common errors occur when interpreting a calibration certificate?

    The most frequent errors are reading “calibrated” as “within tolerance”, ignoring the uncertainty, applying corrections with the wrong sign and using the instrument outside the points or conditions the certificate actually covers.