A standard is a physical object or an instrument, and as such it is subject to the same processes of change as any other. The difference is that its users rely on its value to calibrate everything else, so undetected deterioration spreads to everything calibrated with it. The specific mechanism depends on the type of standard.
Mechanisms by type
- Dimensional standards (blocks, rings, gauges): contact wear, scratches, corrosion, damage to the measuring faces, slow dimensional changes in the material.
- Standard masses: buildup of dirt or moisture, corrosion, impacts that remove material, improper cleaning.
- Electrical standards: component aging, drift of references, contact contamination, overload damage.
- Pressure standards: fluid contamination, wear of pistons and cylinders, fatigue of sensing elements.
- Temperature standards: mechanical stress in sensors, contamination of the sensing element, exposure to extreme temperatures.
- Reference materials: chemical or physical degradation, contamination, expiration.
Causes common to all
- Handling without protection: touching surfaces with bare hands, dropping, knocking.
- Storage without control of humidity, dust or temperature.
- Use outside the conditions for which it was calibrated.
- Intensive use without intermediate checks.
- Transport with vibration or impacts.
How it is detected
Periodic calibration gives a snapshot of its state, but between calibrations it is advisable to compare the standard with a check standard or a more stable reference standard, and to keep control charts with its history. A sustained trend indicates drift; a sudden jump, a damage event.
What to check first
Faced with an unexpected result when calibrating other instruments, check the standard before blaming the instrument. If the standard has deteriorated, it must be withdrawn and its effect on the calibrations performed since its last reliable state must be assessed.