It is to resistance what a standard mass is to mass. Its value, certified with uncertainty, stays stable over time and varies little with temperature, humidity and current. It is calibrated against higher-order standards and is used to verify meters or to establish the resistance scale in a laboratory.
How it is built
Standard resistors are made from metal alloys with stable resistivity and a low temperature coefficient, wound in a way that reduces inductance effects and mechanical stress. They are usually encapsulated, sometimes hermetically or in oil, to protect them from humidity and sudden changes. Many have four terminals: two to pass the current and two to measure the voltage.
Why four terminals
With two terminals, the resistance of the leads and contacts adds to that of the resistor, which is critical at low values. With four terminals, voltage is measured at points separate from those where current is injected, and the effect of the leads practically disappears. The same idea is used in four-wire resistance measurement.
Factors that affect its value
- Temperature: the certified value refers to one temperature; away from it, the coefficient is applied.
- Power dissipation: high currents heat the resistor and change its value; moderate measurement currents are used.
- Drift over time: studied through successive calibrations.
- Humidity and mechanical stress: these mainly affect unsealed constructions.
What the engineer sees in practice
When calibrating a bridge or multimeter, the standard resistor is connected and the reading is compared with its certified value, corrected for temperature if applicable. Letting it reach thermal equilibrium and handling it as little as possible avoids errors. A standard resistor that is damaged or lacks a current certificate is no longer a reliable reference.