A calibration is not a permanent guarantee. It describes how the instrument behaved at certain points, with a certain method and under certain conditions. If any of that changes, the reading can move away from reality even though the certificate is still current. The key is to determine whether the problem lies in the instrument, in the way it is used or in what is being measured.
Causes related to the instrument
- Drift: a slow change in response since calibration.
- Damage or overload: an impact, overpressure or electrical overload can alter the sensor without any obvious sign.
- Wear or contamination: dirt, corrosion, deposits or fatigue of mechanical elements.
- Intermittent faults: loose connections, weak batteries.
Causes related to use
- Outside the calibrated scope: the calibration covers certain points and a certain range; outside them there is no backing.
- Different conditions: temperature, humidity, position or fluid different from those at calibration.
- Poor technique: bad contact, insufficient stabilization time, parallax, varying direction of approach.
Causes related to installation
The instrument may be fine while the measuring point does not represent what you want to measure: a temperature sensor with poor immersion, a pressure tap with trapped air, a connection with a voltage drop, electrical noise in the wiring. This is a very common cause, and a laboratory calibration does not detect it.
What to check first
- Compare against a control instrument or a check standard at the point of use.
- Check ambient conditions, connections and physical condition.
- Confirm that the value is within the calibrated range.
- Repeat the measurement with the correct method and observe the repeatability.
How to interpret the result
If the instrument agrees with the check standard, the problem probably lies in the installation or the process. If it does not agree, take it out of service and assess the impact on previous measurements.