Electronic instrumentation

Clear explanations of concepts, methods, standards and measurements for engineers and technicians.

  • Do thermoelectric voltages cause errors in low-level measurements?

    Yes. Wherever dissimilar metals join and there is a temperature difference, a small voltage appears through the Seebeck effect and adds to the signal, and in low-level measurements it can be comparable to or larger than what you want to measure.

  • EMC requirements set by IEC 61326-1 for laboratory equipment

    IEC 61326-1 sets electromagnetic compatibility requirements for electrical equipment for measurement, control and laboratory use: it limits the emissions the equipment produces and requires a certain immunity to external disturbances.

  • Function generator or arbitrary waveform generator

    A function generator is enough if the test uses standard waveforms such as sine, square or triangle; an arbitrary generator is needed when you must reproduce a captured signal, a special pulse or a specific sequence.

  • Handheld or benchtop multimeter: which to choose

    A handheld multimeter suits field diagnostics, mobility and safety in electrical installations; a benchtop unit suits laboratory measurements where resolution, stability and repeatability matter.

  • How are slow variables logged over several days?

    They are logged with a data acquisition system or recorder set to a low sampling rate, with suitable filtering, a reliable timestamp, backed-up storage and an uninterruptible power supply, verifying the setup before leaving it unattended.

  • How do you build an automated test bench?

    An automated test bench is built by defining what is measured and against what criterion, choosing instruments with a programmable interface, connecting them to the unit under test through a fixture and switching, and controlling it all with software that runs the sequence and records the results.

  • How do you choose the right measurement range?

    The right range is the smallest one that contains the signal without saturating, because an instrument's resolution and accuracy are specified relative to the range and improve when most of it is used.

  • How do you integrate an instrument with remote commands?

    An instrument is integrated with remote commands by sending it text instructions over its communication interface, normally in the SCPI language through a layer such as VISA, and synchronizing each command with the completion of the action.

  • How do you measure a very high resistance without leakage?

    To measure a very high resistance without leakage, you must keep stray currents through insulators, surfaces and cables from flowing through the instrument, using guarding, shielding and quality insulators, and waiting for settling.

  • How do you measure a very low resistance?

    To measure a very low resistance, use a four-wire connection, a test current that is sufficient but controlled, and compensation of thermoelectric voltages, because at these values the wiring and contacts weigh as much as the component.

  • How do you measure a very small current?

    A very small current is measured with a feedback picoammeter or electrometer, or with a very high-resolution multimeter on its lowest range, taking care with leakage, noise and the voltage drop the instrument introduces.

  • How do you measure frequency with a universal counter?

    To measure frequency with a universal counter, connect the signal to the appropriate input, set the coupling, impedance and trigger level, and choose a gate time that gives the desired resolution, with a stable time base.

  • How do you measure the capacitance of an in-circuit component?

    To measure the capacitance of a component in circuit, you must discharge and de-energize the circuit and keep in mind that the other components connected in parallel also contribute to the measured value.

  • How do you measure the regulation of a power supply?

    The regulation of a power supply is measured as the change in its output voltage when the load (load regulation) or the input voltage (line regulation) varies, always measuring at the output terminals.

  • How do you measure the time interval between two events?

    To measure the time interval between two events, connect each event to a channel of a counter or an oscilloscope, choose the edge and trigger level for each one, and measure the difference between the trigger instants.

  • How do you measure the true RMS value of a distorted signal?

    To measure the true RMS value of a distorted signal, you need a true RMS instrument with sufficient bandwidth and crest factor for the waveform.

  • How do you reduce noise in a low-level signal measurement?

    To reduce noise in a low-level signal measurement, you work on three fronts: keeping noise from getting in, lowering the measurement bandwidth, and averaging whatever gets through anyway.

  • How do you verify a multimeter between calibrations?

    A multimeter is verified between calibrations by using it to measure known stable standards, such as a voltage reference and precision resistors, and comparing against limits defined in advance, without this replacing calibration.

  • How does a digital multimeter work?

    A digital multimeter works by conditioning the input quantity into a DC voltage within the range of an analog-to-digital converter, which turns it into a number shown on the display.

  • How does a function generator work?

    A function generator produces periodic electrical signals of known shape, such as sine, square or triangle, with adjustable frequency and amplitude, to stimulate circuits and observe their response.

  • How is a battery characterized with an electronic load?

    The battery is discharged with an electronic load at controlled current, power or resistance while voltage, current and time are logged, down to a defined voltage limit; from this you obtain the capacity, the discharge curve and the internal resistance.

  • How is a start-up test performed on electronic equipment?

    The equipment is powered under defined conditions, its power-up sequence is observed with high-speed recording instruments, and inrush current, internal voltages, timing and final state are verified, repeating under various conditions.

  • How is a switching power supply tested?

    It is tested by operating it under controlled input and load conditions and measuring its output: regulation, ripple, transient response, efficiency, protections and start-up behavior.

  • How is an analog sensor validated on the bench?

    Known, traceable stimuli are applied to the sensor across its full range, its output is measured with a more accurate instrument and compared with the expected response to determine error, linearity, hysteresis, repeatability and stability.

  • Linear or switching power supply for testing

    For noise-sensitive tests, such as analog or low-level circuits, a linear supply is the better choice; to deliver more power in less space with better efficiency, a switching supply, provided its noise is tolerable.

  • Modular data acquisition system or standalone instruments

    A modular data acquisition system is the better choice when you need many synchronized, scalable channels under a single software package; standalone instruments, when you need maximum performance in a specific function or flexibility of use on the bench.

  • Two-wire or four-wire measurement

    Two-wire measurement includes the resistance of the leads and contacts in the result; four-wire measurement eliminates it by separating the conductors that carry the current from those that measure the voltage. It matters above all with low resistances.

  • What causes an instrument’s reading to drift as it warms up?

    Thermal drift occurs because the instrument's components, such as the voltage reference, resistors and amplifiers, change their properties with temperature, and on power-up the instrument goes from a cold state to internal thermal equilibrium.

  • What do half digits mean on a multimeter?

    A rating in digits and a half indicates display resolution: full digits can show zero to nine, and the half digit takes only limited values, normally zero or one. It does not indicate the instrument's accuracy.

  • What does CAT mean in a measurement category?

    CAT is an instrument's measurement overvoltage category: it indicates in which part of an electrical installation the instrument can be safely connected, based on the available energy and the transient overvoltage it must withstand.