ISO 10012 deals with how to manage measurement processes and measuring equipment within an organization, so that measurements are suited to what is to be achieved. It is not limited to calibration: it calls for demonstrating that the measurement system is capable of meeting product or process requirements. Only its general scope is described here; the precise requirements are in the current edition.
Central ideas
- Metrological requirements: it starts from what the customer, the process or the regulations need to measure, and with what quality.
- Metrological confirmation: the set of operations that ensure measuring equipment meets the requirements of its intended use. It includes calibration, verification and, where needed, adjustment or repair.
- Control of measurement processes: design, validation, identification of influence quantities, uncertainty estimation and records.
- Resource management: competent personnel, procedures, software and environmental conditions.
- Improvement: audits, analysis of nonconformities and corrective actions.
Metrological confirmation, not just calibration
A calibrated instrument is not necessarily suitable. Confirmation compares the calibration result with the process requirement: if the error and uncertainty leave margin against the tolerance, the equipment is confirmed for that use. If not, it must be adjusted, moved to another application or replaced. This distinction is the main contribution of the standard.
What the engineer sees in practice
An organization with this system keeps a register of its equipment, the requirement for each item, the confirmation interval, the results and the actions taken, as well as a procedure for evaluating the impact when equipment is found out of tolerance.
When it is used
It is an especially useful guide for organizations that are not laboratories but depend on their measurements, such as manufacturers. It complements other management systems; it does not replace them. Before applying a specific requirement, the current edition must be consulted.