What is a working standard?

A working standard is the standard used routinely to calibrate or verify instruments, and which is in turn calibrated against a higher-level standard to maintain traceability.

Definition

In the hierarchy of standards, the working standard is the one used day to day. It is calibrated with a reference standard, which in turn is calibrated with higher-level standards, up to the national standards and the definition of the unit. That unbroken chain of comparisons, each with its uncertainty, is traceability.

Why there are levels

The most accurate standards are expensive, delicate and valuable. If they were used in every calibration, they would wear out or be damaged. Working standards take on the frequent use, so that the reference standard is reserved for calibrating the working standards. At each step in the chain some accuracy is lost, because uncertainty accumulates.

Essential requirement

A working standard must be appreciably better than the instrument it calibrates. If the uncertainty of the standard is comparable with the tolerance of the instrument, it is not possible to conclude whether the instrument complies. That is why a favorable ratio between the two is sought.

Examples of working standards

  • A high-accuracy multimeter used to calibrate electrical field equipment.
  • A calibrated resistance thermometer used to compare sensors in a bath.
  • A set of gauge blocks used to verify dimensional gauges.
  • A reference digital pressure gauge used as a comparator on a test pump.

How it is cared for

  • It is kept in a controlled environment and protected from impacts, overloads and contamination.
  • It is calibrated as often as its stability requires, and its history is reviewed to detect drift.
  • It is verified between calibrations against a check standard.
  • It is used only for what it was designed for and within its range.

What the engineer sees

When a working standard drifts without anyone noticing, all the calibrations made with it inherit the error. That is why controlling it is more critical than controlling an ordinary plant instrument: a single standard influences many results.