The certificate documents a past fact: the state of the instrument on the date of calibration. Its validity depends on whether that state is still considered representative. That is why validity is lost through events, not only through the passage of time, and the usual period is set by the user based on risk and history, or is required by a standard or a customer.
Situations in which it stops being reliable
- Expired interval: the defined time passed without recalibration, so there is no longer evidence that the instrument remains within tolerance.
- Impact, overload or failure: any event that could have altered the instrument.
- Repair or modification: replacing components changes the behavior.
- Later adjustment: if someone adjusts the instrument after calibration, the data in the certificate no longer describe its state.
- Change of use: working outside the range, or under very different conditions.
- Suspect result: an intermediate check that shows a deviation.
- Broken seals or marks: if the instrument had them to protect the adjustment.
When the problem is in the certificate
It may also be that the certificate never met what was needed: it does not identify the instrument unambiguously, does not show the traceability of the standards, does not state the uncertainty, or does not cover the range and points of use. In those cases it does not lose validity; it never had any for that purpose.
If the standard loses traceability
If it is later discovered that the standard used was out of calibration or damaged when the instrument was calibrated, the calibration is in doubt and its effect on everything calibrated with that standard must be assessed.
What to do
- Withdraw the instrument or identify it as unreliable.
- Recalibrate it, or verify it with a standard before continuing to use it.
- Assess the impact on the measurements made since the last reliable state.