Oscilloscopes

Clear explanations of concepts, methods, standards and measurements for engineers and technicians.

  • What does the IEEE 1241 standard provide for analog-to-digital converters?

    The IEEE 1241 standard provides terminology and test methods for analog-to-digital converters: it defines what their parameters mean and describes how to measure them, so that the specifications of different converters can be compared.

  • What does the IEEE 181 pulse measurement standard cover?

    The IEEE 181 standard defines the terms and algorithms used to measure transitions, pulses and related waveforms, such as rise time, overshoot or pulse width, so that measurements made on different instruments are comparable.

  • What is a differential probe?

    A differential probe measures the voltage difference between its two inputs and rejects what they have in common, making it possible to measure between two points that are not referenced to the oscilloscope's ground.

  • What is a low-capacitance active probe?

    A low-capacitance active probe uses an input amplifier very close to the tip to present a small input capacitance, which loads the circuit lightly and preserves fast edges up to high bandwidths.

  • What is a measurement cursor?

    A measurement cursor is a line placed on the waveform to read the voltage or time value at that point; with two cursors you also get the difference between them.

  • What is a mixed signal oscilloscope?

    A mixed signal oscilloscope combines the analog channels of an oscilloscope with the digital channels of a logic analyzer, so that all signals are captured with the same time reference and the same trigger.

  • What is a passive probe and when is it not enough?

    A passive probe is a cable with resistive attenuation and capacitive compensation, with no amplifier; it is enough for signals of moderate frequency and amplitude, and stops being enough when loading, bandwidth, reference or voltage level rule it out.

  • What is a sampling oscilloscope?

    A sampling oscilloscope reconstructs very fast repetitive signals by taking a sample at a different point on each repetition, using the equivalent-time sampling method, which is why it reaches bandwidths a real-time oscilloscope cannot.

  • What is aliasing and how is it avoided?

    Aliasing is the error whereby an oscilloscope or data acquisition unit shows a fast signal as if it were a slower one, because the sample rate is insufficient to represent it. It is avoided by sampling faster or by limiting the frequency content of the signal.

  • What is equivalent-time sampling?

    Equivalent-time sampling is a technique that reconstructs a repetitive signal from samples taken over different cycles, each with a different time offset, which gives a much higher time resolution than real-time sampling allows.

  • What is memory depth?

    Memory depth is the number of samples an oscilloscope can store in one capture; it determines how much time can be recorded while maintaining a high sample rate.

  • What is the bandwidth of an oscilloscope?

    The bandwidth of an oscilloscope is the frequency at which a constant-amplitude sine wave is displayed attenuated to about seventy percent of its true value, the minus three decibel point, and it marks how far you can trust the measured amplitude.

  • What is the probe attenuation factor?

    The probe attenuation factor is the ratio between the voltage at the tip and the voltage reaching the oscilloscope input; the instrument uses it to scale the reading and display the actual voltage at the tip, which is why it must match the probe in use.

  • What is the rise time of an instrument?

    The rise time of an instrument is the time its output takes to go from a low reference level to a high one when it receives an ideal step, and it is related to its bandwidth: the greater the bandwidth, the shorter the rise time.

  • What is the time base?

    The time base is the oscilloscope's horizontal setting that sets how much time each screen division represents; it determines the visible time window and, on a digital instrument, the sampling rate used for capture.

  • What is the trigger on an oscilloscope?

    The trigger is the mechanism that decides the instant at which acquisition begins, so that every sweep starts at the same point on the signal; without a stable trigger, a periodic signal would appear shifted and blurred.

  • What is waveform update rate?

    The update rate is the number of complete captures an oscilloscope acquires and displays per second. The higher it is, the less dead time there is between captures and the more likely it is that rare, sporadic events will be seen.

  • What safety requirements does IEC 61010-2-030 set?

    IEC 61010-2-030 is a part of the IEC 61010 safety series that adds particular requirements for equipment with testing and measuring circuits, supplementing the general requirements of part one of the series.

  • Why does a probe ground loop distort a fast signal?

    The probe ground loop adds an inductance in series with the input capacitance; together they form a resonant circuit that distorts fast edges with overshoot and ringing that do not belong to the circuit being measured.

  • Why does a waveform show oscillations that do not exist?

    A waveform shows nonexistent oscillations because of aliasing, interpolation across too few samples, resonance of the probe itself and its ground connection, or interference entering the measurement.

  • Why does mains hum appear on the trace?

    Mains hum appears on the trace because the probe, its ground lead or the circuit itself picks up the installation's field at the line frequency, or because a ground loop drives mains current through the measurement connections.

  • Why does the probe load the circuit under test?

    The probe loads the circuit because it adds its resistance, its capacitance and the inductance of its ground connection in parallel; at low frequency the resistance is what counts, and at high frequency the capacitance dominates.

  • Why is a channel input damaged by overvoltage?

    A channel input is damaged when the voltage, energy or current of an overvoltage exceeds what the input protection can withstand, especially on the low-impedance input and on a direct connection without an attenuating probe.