Definition
The time base is expressed in units of time per division. If the screen has a fixed number of horizontal divisions, the total visible time is the product of that number and the chosen value. When you reduce the time per division, you see a shorter interval in more detail; when you increase it, you see a longer interval in less detail.
What changes besides the window
On a digital oscilloscope, memory is finite. The record length, divided by the total visible time, determines the sampling rate. With a fast time base, the instrument can use its maximum sampling rate. With a slow one, to fill the window with the same amount of memory it must take samples further apart, and the sampling rate falls. The deeper the memory, the later this drop occurs.
Practical consequences
- Aliasing: with a slow time base, the sampling rate may be insufficient for the signal and show a false frequency.
- Edge detail: to measure a rise time you need a time base that puts several samples within the edge.
- Rare events: a long window covers more time but with less detail; a short window may miss the event.
- Slow sweep mode: for very long times, some instruments switch to a mode in which the trace is drawn progressively, with no pretrigger.
How to choose it
- Calculate the period of the signal and choose a time base that shows a few periods for an overview.
- Zoom in to a faster time base to study an edge or a detail, using the horizontal delay to bring it to the center.
- Check the sampling rate being used, which is usually shown on the screen.
How to interpret the screen
Before measuring an interval, confirm the horizontal scale. A common mistake is reading the time per division of a different capture from the one being viewed. And if the waveform changes appearance when you change the time base, aliasing is probably present.