Measuring means disturbing
An instrument cannot observe without touching. To obtain the signal, the probe draws a small current from the node being measured. If the source impedance is low compared with that of the probe, this current is negligible. If the source has a high impedance, the presence of the probe changes what you want to measure.
The three components of loading
- Input resistance: forms a divider with the source impedance and reduces the voltage at DC and low frequency. It matters at high-impedance nodes, such as a divider of large resistors or a sensor output.
- Input capacitance: forms a low-pass filter with the source impedance. It slows edges, attenuates high-frequency components and can shift the frequency of an oscillator or destabilize an amplifier.
- Inductance: that of the ground lead and the tip resonates with the input capacitance and adds ringing on fast signals.
What the engineer sees
Slower edges when the probe is connected, a circuit that stops oscillating or oscillates differently when touched, a DC level that drops, or a logic error that disappears when measured. This last clue is the most revealing: if the fault hides when you connect the probe, the probe’s capacitance is modifying the circuit’s behavior.
How to reduce it
- Use an attenuating probe, which increases the resistance and reduces the capacitance seen from the tip.
- For high frequencies, use a low-capacitance active probe.
- Measure at a low-impedance point whenever the circuit allows it, such as the output of an amplifier rather than the high-impedance node before it.
- Shorten the tip and the ground with the appropriate accessories.
- If the effect is still large, compare the measurement with a simulation or with a calculation of the divider formed by the source and the probe.
When it does not apply
For low-impedance nodes and moderate frequencies, the loading of an attenuating probe is negligible, and any differences you observe come from other causes, such as the ground connection or the bandwidth.