How it works
The most common passive probe attenuates the signal with a divider formed by a resistor in the tip and the oscilloscope’s input resistance. An adjustable capacitor compensates for the capacitance of the cable and the input so that the divider attenuates all frequencies equally. On connecting it, that compensation is adjusted with the oscilloscope’s square-wave calibration signal until the edges are flat, with no overshoot or rounding.
Advantages
- Rugged, inexpensive and requires no power.
- High input resistance, so it does not load most circuits at DC.
- Greater voltage range than the instrument’s direct input, thanks to the attenuation.
- Bandwidth suitable for a wide variety of general applications.
When it is not enough
- Fast or high-impedance signals: the probe’s input capacitance, however small, forms a filter with the source impedance, slows the edges and can alter the circuit. In that case, a low-capacitance active probe is advisable.
- Small signals: attenuation reduces the signal, and the probe’s own noise weighs more heavily. A unity-ratio probe or a coaxial cable may be suitable.
- Floating voltages: the passive probe’s reference is the oscilloscope’s ground. If the circuit’s reference point is at a different voltage, you need a differential probe or an isolated instrument, and you must never disconnect the equipment’s protective ground.
- Current: measuring current requires current probes or sense resistors.
- Very high voltage: there are high-voltage probes with suitable insulation and ratings, and an ordinary passive probe must not be used above its limit.
What to check first
- That the compensation is adjusted on the channel where you are going to use the probe.
- That the attenuation factor set on the channel matches that of the probe.
- That the ground connection is as short as possible, since its inductance produces oscillations on fast signals.
- That the voltage and bandwidth of the signal are within what the probe specifies.
Measurements on energized circuits or circuits with hazardous voltages must be made by qualified people, following the applicable safety regulations and using accessories with the appropriate category and voltage rating.