What is the bandwidth of an oscilloscope?

The bandwidth of an oscilloscope is the frequency at which a constant-amplitude sine wave is displayed attenuated to about seventy percent of its true value, the minus three decibel point, and it marks how far you can trust the measured amplitude.

Practical definition

An oscilloscope behaves like a low-pass filter at its analog input. Bandwidth is the frequency at which that filter attenuates the signal by three decibels, which is equivalent to displaying about seventy percent of the amplitude. It is not a sharp limit: attenuation starts earlier, and appreciable amplitude error already appears below that frequency.

Relationship with rise time

A system with limited bandwidth cannot produce arbitrarily fast edges. The instrument’s rise time is inversely proportional to bandwidth, with a constant that depends on the shape of the response. When the signal’s edge is comparable to that of the instrument, the measured edge is rounded and looks slower than it really is. The observed rise time combines those of the signal, the probe and the oscilloscope, approximately as a root sum of squares, so every link in the chain counts.

Why a digital signal demands more than its fundamental frequency

A square wave contains odd harmonics of its fundamental, and fast edges depend on frequency components much higher than the fundamental. With a bandwidth that only reaches the fundamental, the square wave appears as a rounded sine wave. To see its shape, the bandwidth must cover several harmonics or, better still, be consistent with the rise time of the edges.

The system includes the probe

The oscilloscope’s rated value holds only if the probe also allows it; the bandwidth of the combination is lower than that of each element separately. Probe capacitance and ground lead inductance limit the result and produce ringing on the edges.

What you see in practice

  • Amplitudes lower than expected on fast signals.
  • Rounded edges and rise times that do not change when the circuit changes.
  • Aberrations such as overshoot when approaching the limit.

How to choose it

Start from the fastest signal you need to observe, consider its rise time and the tolerable amplitude error, and add margin. A higher bandwidth also lets more noise through; many instruments include a limiting filter for this purpose.