What it provides
An analog-to-digital converter is not an ideal device: it has gain, offset and linearity errors, as well as noise. The datasheet summarizes them in parameters, and IEEE 1241 establishes how they are defined and how they are measured. Its practical value is that it gives the same meaning to terms that, without it, each manufacturer would use in its own way.
Parameters it organizes
- Static: gain and offset error, differential and integral nonlinearity, and missing codes. They describe how the input range is divided among the output codes.
- Dynamic: signal-to-noise ratio, harmonic distortion, signal-to-noise and distortion ratio, spurious-free dynamic range and effective number of bits. They describe behavior with signals that change over time.
- Timing: aperture uncertainty or sampling jitter, and settling time.
How it is tested, in general terms
For static parameters, known input voltages or slow ramps are applied and the response in codes is analyzed, for example using histograms. For dynamic parameters, a pure sine wave is applied and the data are analyzed with a Fourier transform or a sine fit, separating the signal from the noise and harmonics. This yields the effective number of bits, which is lower than the nominal number.
Relationship to oscilloscopes
The nominal resolution of an oscilloscope is that of the converter, but the effective resolution, which includes noise from the analog input, is what matters on screen. Understanding these concepts helps explain why a small signal looks noisy or why averaging improves the measurement.
What it is not
It does not define what value each parameter should have, nor is it a standard on the complete instrument system. Like all standards, it is updated, so consult the current edition to confirm the terms and procedures.