What happens in each method
In two-wire measurement, the instrument sends current through the same conductors it uses to measure the voltage. The reading is the sum of the resistance of interest, that of both leads and that of the contacts. In four-wire measurement, one pair of conductors injects the current and another pair, connected at the ends of the resistance, measures the voltage. Because the voltmeter has a very high input impedance, almost no current flows through the sense leads and their voltage drop is negligible.
When the difference is noticeable
The lead error is an added resistance, so it carries weight when the measured resistance is small. At a high value, of several kilohms, the lead resistance is insignificant and two wires are enough. On a milliohm resistance, such as a shunt, a soldered joint or a contact, two wires can give a completely wrong result. Dirty or loose contacts add a variable resistance, which shows up as an unstable reading.
What you see in practice
- In two-wire measurement, shorting the probes shows a nonzero value; that residual is the error, which is subtracted if the instrument allows it.
- In four-wire measurement, the value is stable and does not change when the sense leads are lengthened.
- If the instrument has no four-wire function, the same result can be achieved with a current source and a separate voltmeter.
Other effects that remain
Thermoelectric voltages at junctions of dissimilar metals introduce an offset that affects both methods, and it is compensated by reversing the polarity of the current. In addition, the measurement current heats the resistance, which changes its value if the material is temperature-sensitive.
How to decide
Compare the expected value with the resistance of the leads and contacts. If it is comparable or smaller, use four wires. With energized or power circuits, connections are made by qualified personnel in accordance with the applicable regulations.