What is being measured
The effective value, or RMS, of a signal is the value of a DC voltage that would produce the same heating in a resistor. For any signal, it is obtained by squaring the signal, averaging and taking the square root. It is the quantity that matters when power or heating is involved, and it is the same for a distorted waveform as for a sine wave only if the energies match.
Why an ordinary instrument fails
Average-responding instruments rectify the signal, calculate its average and multiply it by a factor that is correct only for a pure sine wave. With distorted waves, such as those from a rectifier load, a variable-speed drive or a switching power supply, that conversion is no longer valid and the error can be large. A true RMS instrument calculates the real quantity, either with a thermal converter or by digital calculation, and does not depend on the waveform.
What to check on the instrument
- Allowable crest factor: the ratio between the peak and RMS values. Signals with narrow pulses have a high crest factor, and the instrument must handle it without saturating. An accuracy specification usually applies up to a certain crest factor.
- Bandwidth: the signal’s harmonics contribute to the RMS value. If the instrument does not reach them, it loses them. The specification is usually worse at high frequencies.
- Coupling: an AC-only mode discards the DC component; an AC+DC mode measures the total. For the RMS value of the complete signal, use the second, or combine the two components in quadrature.
- Range: consider the peaks, not just the expected RMS value.
How to interpret the result
If the same signal is measured with an averaging instrument and a true RMS instrument and the results differ, the signal is distorted and the second one is the reliable one. An oscilloscope with RMS measurement also provides the waveform, so you can see where the difference comes from.
For measurements on power circuits, connections are made by qualified personnel in accordance with electrical safety regulations.