How do you measure a very high resistance without leakage?

To measure a very high resistance without leakage, you must keep stray currents through insulators, surfaces and cables from flowing through the instrument, using guarding, shielding and quality insulators, and waiting for settling.

The problem

Measuring a very high resistance amounts to measuring a tiny current when a known voltage is applied. At these levels, leakage currents through insulators, circuit boards, connectors and damp or dirty surfaces are comparable to the current being measured, and they add to it. The result is an apparent resistance lower than the real one.

How leakage is avoided

  • Guarding: a conductor surrounds the measurement point and is kept at the same potential. With no voltage difference between the two, no leakage current flows to the guarded element; the leakage that does flow is diverted to the source and does not pass through the instrument.
  • Shielding: a properly connected metal enclosure prevents external electric fields from inducing currents in the high-impedance circuit.
  • Low-leakage insulators: materials such as PTFE for supports, and suitable coaxial or triaxial cables. Avoid insulators that absorb moisture.
  • Cleanliness and a dry environment: finger grease and moisture create conductive surface paths.

Measurement method

A known voltage is applied with a source and the resulting current is measured with an electrometer or picoammeter, or an instrument designed for high resistances is used. The resistance is then calculated with Ohm’s law. The voltage must be sufficient to produce a measurable current without exceeding what the sample or the instrument can withstand; in some materials the resistance depends on the applied voltage.

Settling time

When the voltage is applied, a charging current for the capacitances and a dielectric absorption current from the material flow first, and they decay slowly. Measuring too soon gives high currents and low resistances. That is why a defined, constant time is allowed, and it is reported along with the result. Moving cables generate triboelectric charges that appear as noise currents, so they must be secured.

How to interpret the result

A reading that increases over time toward a stable value indicates dielectric absorption. A reading that changes with humidity or when the setup is touched indicates surface leakage. Comparing with and without guarding helps estimate how much was due to leakage.

High test voltages can be hazardous: they must be handled by qualified persons in accordance with the applicable safety regulations.