How do you verify a multimeter between calibrations?

A multimeter is verified between calibrations by using it to measure known stable standards, such as a voltage reference and precision resistors, and comparing against limits defined in advance, without this replacing calibration.

Verifying is not calibrating

A calibration compares the instrument against traceable standards and documents its deviation and uncertainty. An intermediate verification is a simpler, more frequent check that seeks to detect whether the multimeter has changed appreciably since its last calibration. It does not adjust anything or issue a certificate: it only gives confidence, or a warning.

What you need

  • Stable check standards: a DC voltage reference, precision resistors of stable values and, if required, a current source or a calibrator. Their stability must be better than that of the multimeter being monitored.
  • Representative points: one or more values on the functions and ranges actually used, and also the zero with the input shorted.
  • Controlled conditions: stable temperature, instrument warm-up, clean cables and connections.

How it is done

  1. Let the instrument warm up.
  2. Measure the zero with a short on the voltage and resistance functions, using lead compensation if four wires are used.
  3. Measure the standards at the chosen points and repeat several readings.
  4. Compare with the value of the standard and with the acceptance limit.
  5. Record the date, temperature, value and result.

Criteria and trends

The acceptance limit must be derived from the multimeter’s specification and the uncertainty of the check standard, and set before measuring. More useful than an isolated result is the trend: if the values shift consistently in one direction, the instrument is drifting and should be reviewed before it goes out of tolerance. A control chart of the records shows this well.

How to interpret the result

A result within limits does not prove that the instrument is calibrated on all its ranges; it only indicates that it has not changed at the points checked. If a verification fails, the instrument is taken out of service, the measurements made since the last good verification are reviewed, and it is sent for calibration.

Common mistakes

  • Using as a check standard an instrument less stable than the one being monitored.
  • Verifying without waiting for warm-up.
  • Not keeping the history, so drift goes undetected.
  • Relying on a single function when the multimeter is used on several.