EMC requirements set by IEC 61326-1 for laboratory equipment

IEC 61326-1 sets electromagnetic compatibility requirements for electrical equipment for measurement, control and laboratory use: it limits the emissions the equipment produces and requires a certain immunity to external disturbances.

Two sides of electromagnetic compatibility

Electromagnetic compatibility involves two complementary requirements: that the equipment does not disturb other equipment beyond reasonable limits, that is, emissions, and that it keeps working properly in the face of disturbances from its environment, that is, immunity. The standard applies both to measurement, control and laboratory equipment.

Emissions

Disturbances conducted along power and signal cables and those radiated into the air are assessed. The limits depend on the classification of the environment and the equipment, which distinguishes, for example, residential from industrial environments. Typical sources are digital clocks, switching power supplies and power switching.

Immunity

The equipment is subjected to representative phenomena: electrostatic discharge, radiated radio-frequency fields, electrical fast transient bursts, surges, conducted radio-frequency disturbances, and voltage dips or interruptions. Test levels depend on the intended environment.

Performance criteria

Not every effect counts as a failure. The standard distinguishes how the equipment may behave during and after the disturbance: normal operation, temporary degradation that recovers on its own, or loss of function that requires operator intervention. For a measuring instrument, a certain degradation of accuracy is sometimes allowed during the test, provided it is declared and the instrument returns to normal.

What the engineer sees in practice

  • Noisy or offset readings near welding equipment, a variable-speed drive or a radio.
  • Instrument restarts after an electrostatic discharge.
  • Long measurement cables that act as antennas and pick up disturbances.

Good wiring, shielding and grounding practices reduce these effects, and manufacturers state the conditions under which they met the standard, such as the use of shielded cables.

When it does not apply

An instrument’s conformity does not guarantee correct results in every installation. For exact scope and limits, consult the current edition of the standard.