How Do You Choose a Standard to Calibrate an Instrument?

A standard is chosen so that its uncertainty is clearly smaller than that of the instrument being calibrated, it covers the required range and quantity, it is compatible with the method and it has a current traceable calibration.

The central rule is that the standard must be better than the instrument being calibrated. If both have similar uncertainties, the calibration result cannot tell whether the deviation comes from the instrument or from the standard. This is why a ratio of at least several times between the instrument requirement and the uncertainty of the standard is usually sought, although the specific ratio depends on the criteria of each organization.

Selection criteria

  • Uncertainty: that of the standard, including its stability since the last calibration, must be small compared with the instrument’s error limit.
  • Range and points: the standard must cover the entire span over which the instrument is used.
  • Resolution: that of the standard must be better than that of the instrument, so as not to limit the reading.
  • Traceability: a current certificate that identifies its chain to the reference.
  • Compatibility: same type of signal, connection, medium or fluid, depending on the quantity.
  • Stability and condition: undamaged, with a known drift history.

What the engineer sees in practice

A typical mistake is calibrating a precision pressure gauge with another of the same type and class. The readings agree, but the calibration provides no useful information because the uncertainty of the standard consumes the entire margin. Another case is using a standard whose certificate covers only part of the range: outside it, traceability is not supported.

Conditions of use of the standard

The uncertainty on the certificate is valid for the conditions under which the standard was calibrated. If it is used at a different temperature, orientation or with a different fluid, there are effects to add to the uncertainty budget. It is also advisable to verify the standard between calibrations with an intermediate check.

What to check first

Before using a standard, confirm that its certificate is current, that the range covers the job and that the stated uncertainty leaves enough margin. If any condition fails, it is better to change standards than to accept a calibration that proves nothing.