Indication error is the difference between what the instrument shows and the conventional true value, which in practice is the value given by a traceable standard. It is obtained by applying known values of the quantity to the instrument, reading its indication and subtracting.
General procedure
- Preparation: the instrument and the standard are stabilized under suitable environmental conditions, and zero, connection and cleanliness are checked.
- Point selection: points are chosen across the span, including the values at which the instrument is used most.
- Applying the value: the same value is generated or measured with the standard and with the instrument, waiting for stabilization.
- Reading: both readings are recorded.
- Calculation: error equals indication minus reference, at each point.
Details that change the result
- Direction of approach: going through the points upscale and downscale reveals hysteresis.
- Repeats: several series give an estimate of repeatability.
- Correction of the standard: if the standard’s certificate gives corrections, they are applied to the reference value.
- Environmental effects: temperature and other conditions can alter the reading.
How to interpret the result
The error alone does not tell you whether the instrument is fit for use. It must be compared with the permitted limit, taking uncertainty into account, and the trend across the span must be observed: a constant error suggests a zero problem; one that grows with the value, a gain problem; an irregular one, a problem of linearity or of the instrument’s condition.
When this scheme does not apply
Some instruments, such as those that measure by direct comparison between two quantities, or those that cannot be covered at discrete points, require adapted procedures. The idea of comparing against a traceable reference still holds.