Why preparation matters
Calibration measures the state of the instrument as it is used. If it arrives dirty, damaged, without accessories or freshly adjusted, the result no longer represents its performance in operation, or the calibration is delayed by having to correct what could have been resolved beforehand.
What to do beforehand
- Inspect visually. Look for dents, damaged threads, broken glass, stripped cables, corrosion or loose connectors. Any visible damage must be reported before calibration.
- Clean without altering. Remove process residue, oil or dust, using a method suited to the material. On pressure equipment, keep incompatible fluids from contaminating the calibrator’s standard.
- Gather the accessories. Probes, cables, adapters, power supplies, charged batteries, manuals and software are part of the system being calibrated.
- Document the use. Actual working range, units, critical points, required tolerance, history and operating conditions, so that the correct points are chosen.
- Acclimatize. An instrument coming from a very different environment needs time to match the calibration conditions, especially temperature and humidity.
- Confirm basic operation. Check that it powers on, shows no errors and has a reasonable zero.
What not to do
Do not adjust the instrument before calibrating it, even if you suspect it has deviated. The “as-found” data are what make it possible to know whether previous measurements were valid. Likewise, do not change settings, filters or scales without recording it.
An example
A pressure transmitter arrives without information on its configured range or the fluid it operates in. The calibrator may verify it in a range different from the one used in the plant, and the certificate will not demonstrate its fitness for the process. With the range and points of interest stated from the start, the calibration concentrates where it matters.
Safety
Equipment that has been in service with pressure, hazardous products or voltage must be depressurized, decontaminated and de-energized by qualified personnel following the applicable safety procedures before it is handled or shipped.