They are not the same: calibration compares the instrument with a standard and documents its error; adjustment modifies the instrument to reduce that error. You can calibrate without adjusting, and adjust without having validly calibrated.

To calibrate is to measure and document the relationship between the instrument’s indications and the values of a standard, with their uncertainty. It does not change the instrument. To adjust is to intervene in it, through zero, gain, coefficients or components, so that its indications move closer to the reference values. The confusion is common because in the shop both are usually done during the same visit.

What each one produces

  • Calibration: a set of errors or corrections with uncertainty and traceability.
  • Adjustment: an instrument with an error state different from the one it had.

Why it matters to keep them separate

If the instrument is only adjusted and the final state is reported, there is no record of how far it had deviated. That information is what shows whether measurements made since the previous calibration may have been incorrect, and it is what allows drift to be evaluated. Good practice is therefore to calibrate first and record the as-found state, adjust if appropriate, and then calibrate again to record the as-left state.

When to adjust and when not to

Adjustment is not mandatory. If the error found falls within the acceptance criterion and is stable, it may be preferable to leave it and apply, if needed, a documented correction. Adjusting blindly introduces variation and erases evidence. On the other hand, if the error exceeds the criterion, adjustment is the usual course, provided the instrument has the means for it and the deviation does not reflect damage.

Special cases

  • An instrument without adjustment controls cannot be adjusted; it can only be calibrated and corrections applied.
  • In some regulated instruments, adjustment requires seals or authorization from the manufacturer.
  • After a repair, adjustment and subsequent calibration are mandatory before returning to service.

An example

A balance shows a small, constant deviation across the entire span. The calibration documents it. If the deviation exceeds the acceptable limit, the balance is adjusted with a standard mass and the calibration is repeated to confirm the result. A certificate showing both states makes it possible to understand what happened.