How Is an Adjustment Performed After a Calibration?

An adjustment after calibration is made by modifying the instrument so that its indication moves closer to the reference value. It must always be preceded by recording the error found and followed by a verification of the result.

Adjustment is an operation on the instrument, not part of calibration in the strict sense. Calibration compares and documents; adjustment changes the instrument to reduce its error. The correct sequence is therefore: calibrate and record the as-found state, decide whether adjustment is needed, adjust, and calibrate or verify again to record the as-left state.

Work sequence

  1. As-found state: the points are measured with the instrument as it arrived and the errors are recorded. These data are the basis for evaluating drift and the impact on previous measurements.
  2. Decision: the result is compared with the acceptance criterion; if the error is acceptable and stable, it may not need adjusting.
  3. Adjustment: action is taken using the means the instrument provides.
  4. As-left state: the instrument is measured again to confirm that the error is now within the criterion.

What is adjusted

Many instruments have two basic controls: zero, which corrects a constant offset, and gain, or span, which corrects proportional error. It is best to adjust zero first and gain afterward, and to repeat if the controls interact. Digital instruments may have correction tables or linearization coefficients instead of potentiometers. The manufacturer’s procedures must always be followed.

When to avoid adjusting

Adjusting without need can be counterproductive: the drift history is lost and variation is introduced. If the error is small compared with the tolerance, it is sometimes preferable to leave it and apply a documented correction. Nonlinear error or error with hysteresis cannot be resolved with zero and gain either; it indicates an instrument problem that may require repair.

What the engineer sees in practice

A well-prepared certificate shows data before and after adjustment. If only the final state appears, it is impossible to know whether the instrument had drifted, and there is no basis for deciding the calibration interval.