How each one works
In an analog oscilloscope, the amplified signal directly deflects an electron beam in a cathode-ray tube. The trace is the signal itself, it is drawn in real time, and its brightness depends on how many times the beam passes over a point. In a digital one, an analog-to-digital converter takes samples and stores them in memory, and a processor draws them on the screen. Between the signal and what you see there is a reconstruction.
What digital offers
- Capture of single events and storage, with the ability to freeze, save and compare.
- Pretrigger: you see what happened before the event, because the instrument records continuously.
- Automatic measurements, cursors, FFT, decoding and math on the waveforms.
- Large memories and later analysis.
What analog retains
- A display with no processing delay and a very high update rate.
- Natural intensity grading: infrequent events look faint and repeated ones look bright. Many digital oscilloscopes imitate this with persistence or color-graded displays.
- No aliasing, since there is no sampling.
Limitations of digital
If the sample rate is insufficient for the signal, aliasing appears: a false frequency lower than the real one. Events that occur between captures can be lost if the dead time between acquisitions is long. And interpolation between samples can insert oscillations the signal does not have. The limitations of analog are different: it does not store, it does not show what precedes the trigger, and its bandwidth and calibration depend on aging components.
How to decide
For almost any modern diagnostic work, a digital instrument is the choice. An analog one is justified in a teaching laboratory to see the direct relationship between signal and trace, or when one is already on hand and its bandwidth is sufficient. A digital instrument with a high waveform update rate covers much of the advantage of analog.