Definitions
The range is the interval of values of the quantity for which the instrument is calibrated or specified, from the lower limit to the upper limit. The span is the algebraic difference between those limits. An instrument calibrated from 0 to 10 units has a span of 10; one calibrated from 2 to 12 also has a span of 10 but a different range.
Why the distinction is made
Two instruments with the same span can start at different points on the scale, and two with the same starting point can have different spans. When an instrument is adjusted, the zero shifts the whole range and the span changes the slope between input and output. These are two independent adjustments that are interpreted separately.
Sensor range and calibrated range
A transmitter has a limit range determined by the sensor and a configured range that the user chooses within that limit. Reducing the configured range relative to the limit is commonly called turndown; it can improve resolution, but it normally does not improve accuracy in the same proportion, and the behavior depends on the manufacturer’s model.
How it affects accuracy
Many specifications express error as a percentage of span or of full scale. On that basis, the permitted absolute error is constant across the whole scale; therefore, at low values the error relative to the reading is greater. For this reason the working value should fall in the middle or upper part of the range, not at the lower end.
An example
A transmitter configured at 4 to 20 mA for a pressure interval has a span of 16 mA at its electrical output and a span in pressure units equal to the difference between the upper and lower limits. The two expressions of span are equivalent, but they apply to different sides of the instrument.
What to check
When reading a data sheet, confirm whether the accuracy refers to span, reading or full scale, and whether the stated range is the sensor limit or the configured range.