What does ISO 9227 cover on salt spray?

ISO 9227 specifies the equipment, reagents and conditions for corrosion tests in an artificial salt spray atmosphere, used to evaluate metals and coatings.

General scope

ISO 9227 describes corrosion tests performed in an artificial salt spray atmosphere. It defines the apparatus, reagents and procedure for evaluating the corrosion resistance of metals with or without protection, and of metallic, conversion and organic coatings. It applies to samples whose behavior is to be compared or controlled.

Which variants it includes

The standard provides for different solutions and conditions. The best known is neutral salt spray, with sodium chloride. Other variants use acidification with acetic acid, and one more also adds copper chloride to accelerate the attack on certain decorative finishes. Each one is intended for a type of material and coating. The exact conditions are found in the current edition.

What it establishes

  • The equipment requirements, including the chamber material and the need for the fog not to drip onto the parts.
  • The preparation of the solution and its purity, concentration and acidity.
  • The enclosure temperature and the fog deposition rate, which is measured with collectors.
  • The placement of the samples, their orientation and the spacing between them.
  • The use of reference specimens to check the reproducibility of the equipment.

What it does not establish

The standard describes how to expose the samples, not for how long or when they pass. The duration and acceptance criteria come from the product specification or the coating standard. Nor does it define the evaluation of results, which is governed by other documents.

How to interpret the result

The salt spray test is useful for comparing and controlling quality. It should not be taken as a direct predictor of service life, because the correlation with real corrosion varies with the material and the environment. For an approximation of service conditions, cyclic tests described in other documents are used.

How to consult it

Before planning, verify the current edition and confirm which variant the specification requires. It is common for two laboratories to disagree simply because they use different variants or editions.