Definition
A high-resolution multimeter does not take an instantaneous sample. It integrates the signal over a period of time, essentially summing or averaging its value, and converts the result. The longer that interval, the more it averages, so random, uncorrelated noise decreases relative to the signal.
Mains frequency rejection
The most common disturbance in a measurement is the one at the mains frequency. If the integration time is an integer multiple of the mains period, the integral of a sine wave at that frequency is zero and the interference cancels out. That is why many instruments express this time in power line cycles. If the integration period does not span complete cycles, cancellation is partial.
The inevitable trade-off
- Short time: fast readings, suitable for changing signals, but with more noise and less effective resolution.
- Long time: stable readings and more useful digits, suitable for constant signals, but slow and less suited to varying signals, because the average smooths them.
What you see in practice
As integration time increases, the spread of readings decreases and the instrument shows more stable digits. If the signal varies during integration, the reading corresponds to the average over that interval and not to an instant. In a sweep or a multichannel system, a long time lengthens the complete measurement cycle and must be included when calculating the test time.
How to choose it
- Define how much uncertainty the measurement can accept.
- Check whether mains noise dominates; if so, use whole cycles.
- Compare reading stability at different values and choose the smallest that meets the requirement.
It is a relationship between noise, resolution and speed; there is no universally correct value.