Agilent 3458A
The Agilent 3458A digital multimeter, Keysight 3458A, is a high-speed, high-accuracy instrument for the production test floor, R&D and the calibration laboratory. The HP 3458A was developed by Hewlett-Packard. In your system or on the bench, it saves time and money with its performance and accuracy, its seven-function measurement flexibility and its low cost of ownership.
Select a reading rate of 100,000 readings per second for maximum test throughput, or achieve the highest levels of accuracy with up to 8 1/2 digits of measurement resolution and a transfer accuracy of 0.1 parts per million. It offers programming compatibility through the Hewlett-Packard Multimeter Language (HPML).
Features
Faster testing
- Up to 100,000 readings/s
- Internal test setups > 200/s
- Programmable integration times from 500 ns to 1 s
Higher test throughput
- Better accuracy for tighter test margins
- Resolution of up to 8 1/2 digits
Higher uptime
- Two-source calibration (10 V, 100 kΩ), including AC
- Self-adjusting, self-verifying automatic calibration for all functions and ranges, including AC
Higher waveform resolution and accuracy
- 16 bits at 100,000 samples/s
- 18 bits at 50,000 samples/s
- 12 MHz bandwidth
- Timing resolution to 10 ns
- Timing jitter of less than 100 ps
- More than 75,000 readings of internal memory
Flexible digitizing software
- Analysis software for HP 9000 Series 200/300 computers
- Subprograms for waveform acquisition, data transfer, FFT, IFT and data presentation
Transfer measurements
- 8 1/2-digit resolution
- 0.1 ppm DC volts linearity
- 0.1 ppm DC volts transfer capability
- 0.01 ppm rms internal noise
Accuracy
- 0.6 ppm over 24 hours on DC volts
- 2.2 ppm over 24 hours on ohms
- 100 ppm midband AC volts
- Voltage reference stability of 8 ppm (4 ppm optional) per year