Differential signal and common signal
A differential instrument measures the voltage difference between two terminals. If both terminals also carry the same voltage with respect to ground, that common part is not the signal of interest. It may come from power-line noise coupled into the cables, from different ground potentials between the source and the instrument, or from a sensor bias voltage. An ideal instrument would respond only to the difference.
Definition
Common-mode rejection measures how close the instrument comes to that ideal. It is the ratio of the gain for the differential signal to the gain for the common signal, expressed in decibels. A high value means the common voltage is strongly attenuated before it affects the reading. In practice, the part of the common signal that is not rejected appears at the output as an error, as if it were a small differential signal.
Why it degrades
- Source impedance imbalance: if the two sides have different source resistances, part of the common voltage is converted into differential voltage. The rejection of the real circuit can be worse than the figure in the instrument’s data sheet.
- Frequency: rejection is usually highest at DC and decreases as frequency rises, because the stray capacitances are not identical on both sides.
- Wiring: long cables that are neither twisted nor shielded pick up more common noise.
- Common-mode limits: if the common voltage exceeds the input’s permitted range, the instrument stops behaving correctly.
What the engineer sees
A small signal with a superimposed periodic component at the line frequency, which disappears if the source is shorted but not if the cable is disconnected, points to common-mode noise that is not being rejected. High-resolution multimeters publish specifications for common-mode rejection and for normal-mode rejection, the latter being the ability to ignore noise superimposed on the differential signal.
How to improve the measurement
Use twisted, shielded cables, balance the source impedances, avoid ground loops, and use galvanic isolation when the common voltage is large. With hazardous voltages, connections are the responsibility of qualified personnel following the applicable safety regulations.