What is a TEM cell?

A TEM cell is a flared rectangular transmission line, with a flat center conductor (septum) and metal outer walls, that generates a nearly uniform electromagnetic field inside. It is used to test small equipment for immunity or to measure its emission without a large chamber.

How it is built

A TEM (transverse electromagnetic) cell works as a coaxial line whose cross-section widens to house an object. It has a flat inner conductor called a septum, a closed outer conductor and two cone-shaped ends that taper down to coaxial connectors. The design keeps the characteristic impedance constant along the line, so that the signal travels without significant reflections.

What field it produces

Between the septum and the wall there is an electromagnetic field that approximates a plane wave: an electric field perpendicular to the septum and a magnetic field perpendicular to the electric field. The usable region is in the central part, between the septum and one of the walls.

Uses

  • Immunity: one of the ports is fed with a radio-frequency signal and the other with a matched load. The equipment, inside the cell, is exposed to the known field.
  • Emissions: the equipment under test excites the line and the coupled signal is measured at one port, while the other is terminated in a load. The result is related to the field that would be measured at a test site by means of conversion algorithms.

Advantages and limits

The main advantage is size and containment: a cell requires little space and needs no absorbers, and its field is protected from the ambient environment. Its limits are:

  • The equipment must be small relative to the space between septum and wall, or it distorts the field.
  • Above a certain frequency, which depends on the dimensions, higher-order modes appear and the field is no longer uniform; smaller cells reach higher frequencies, and cells for large equipment work at lower frequencies.
  • The equipment cables leave the cell and alter the correlation with an open-field measurement.

When it does not apply

It is not suitable for large equipment or equipment with many cables of considerable length. For higher frequencies there is the GTEM variant, wedge-shaped and terminated with absorbers, which extends the usable range. In any case, the cell is used as a supporting or comparative evaluation method, depending on what the applicable standard allows.