What is a particle counter used for?

A particle counter measures how many particles are in a liquid and what size they are, normally by passing the fluid through a light beam and recording the signal from each particle that interrupts it; it is used to verify oil cleanliness.

What it is used for

Clean oil is part of the insulation and of the mechanical reliability of equipment with pumps. The counter makes it possible to check cleanliness objectively: after filtering, after filling a new or repaired transformer, or when monitoring wear in equipment with forced circulation.

How it works

The oil passes through a measuring cell crossed by a light beam. Each particle that crosses the beam blocks a fraction of the light and produces a pulse at the detector. The pulse height depends on the size of the particle. The instrument sorts the pulses into size channels and counts how many there are in each one per volume of sample. It is calibrated with a suspension of particles of known sizes.

What an engineer sees in practice

A table of counts by size range and, in many cases, a summary cleanliness code. These are compared with the targets for the equipment. If the counts do not drop after treatment, filters, connections, and hoses must be checked.

What can distort the reading

  • Air bubbles, which are counted as particles; the sample must be degassed or left to stand.
  • Droplets of free water.
  • Oil that is too viscous or very dark, which hinders the passage of light.
  • Settled samples that have not been homogenized.
  • Dirty containers and hoses.

What it cannot tell you

The counter does not know what the particles are made of: cellulose fiber, metal, carbon, and dust are all counted the same way. Identifying the origin calls for analysis by filtration and microscopy.

When it is the right instrument

It is useful for quick cleanliness checks and for tracking over time. The decision to intervene on equipment must be made using other oil indicators and by qualified personnel.