What it simulates
Lightning rarely strikes equipment directly. More commonly, a nearby strike induces voltages on power or communication lines, or the lightning current raises the ground potential. Overvoltages also occur from switching of breakers and capacitor banks on the network. The surge test applies a unidirectional pulse of that class of energy.
Combination wave
The generator is defined by a combination wave: in open circuit it produces a voltage pulse with a front time of 1.2 microseconds and a half-value duration of 50 microseconds; in short circuit it delivers a current pulse of 8 and 20 microseconds. What the equipment sees depends on its impedance: the generator has a defined source impedance, so the pulse resembles a voltage source for high-impedance loads and a current source for a low-impedance one.
Coupling
- Line to line (differential mode): the pulse is applied between live conductors.
- Line to ground (common mode): the pulse is applied between the conductors and the ground reference.
A coupling and decoupling network injects the pulse into the equipment and prevents it from traveling back into the supply network. Communication lines use specific networks with coupling methods suited to the type of line.
What the engineer sees in practice
The failures are more severe than with bursts: protective devices that open, varistors that degrade, fuses that blow, damaged semiconductors or even arcs between traces. Sometimes the equipment survives but resets when the pulse is applied, and the failure appears after several events have accumulated.
How a design is protected
Protection is based on limiting the voltage and diverting the energy: varistors, gas discharge tubes and suppressor diodes, used in coordinated stages. The highest-energy element goes first and the fastest one protects the sensitive part. The path of the diverted current to ground must be short and low in inductance, otherwise the residual voltage on the protected circuit remains high.
These tests involve dangerous voltages and must be carried out only by qualified personnel, in accordance with the applicable regulations.