Basic idea
With two wires, the instrument passes a current through the resistance and measures the voltage across its terminals with the same leads. The measured voltage includes the drop in the leads and contacts, so the result is the sum of the resistance sought plus the resistance of the leads and connections. When the resistance is large this does not matter, but with small resistances the error is of the same order as the measurement.
How the Kelvin method solves it
Two pairs of leads are used. The test current flows through the first pair, called the force pair. The second pair, the sense pair, picks up the voltage directly at the ends of the resistance. Because the voltage measurement circuit has a very high input impedance, almost no current flows through the sense leads, and therefore there is almost no voltage drop in them. The instrument measures the voltage actually present across the component and calculates the resistance from the known current.
Details that matter
- The sense leads must be connected as close as possible to the body of the component, between the force points and the resistance, not before them.
- Kelvin clips and probes physically separate the two contacts at each terminal.
- If the sense point ends up on the wrong side, the contact drop is added back in.
- Thermal electromotive forces at junctions of dissimilar metals also contribute error; they are reduced by offset compensation, for example by reversing the current.
When it is worthwhile
It is used for measuring low resistances, such as windings, shunts, soldered joints, contacts and busbars. It is also used with resistance temperature sensors, where long leads introduce resistance comparable to the variation being measured. When measuring high-value components, the advantage is minimal.
Common mistakes
- Thinking that four leads are enough without taking care over where the sense leads connect.
- Using very long, unshielded sense leads in a noisy environment.
- Passing too much current, which heats the component and changes its value.
- Not checking with a known resistance that the connection eliminates the wiring contribution.