What a climatic chamber does
It reproduces, under control, a thermal environment and, where the chamber allows it, a relative humidity environment. It is used to see how a product behaves under heat, cold or humidity, whether with fixed values or programmed cycles. Its value is repeatability: the same profile can be applied to different samples on different days.
How it works
A fan moves air through a heat exchanger. Electric heating elements provide the heat; cooling comes from a refrigeration system, which at very low temperatures may be a cascade system. Humidity is generated with steam or water evaporation and is reduced by cooling and condensation on a coil. Temperature and humidity sensors feed back to a controller that adjusts heating and refrigeration power.
What distinguishes one chamber from another
- Temperature and humidity range covered.
- Rate of change, important in cycling and thermal shock.
- Temperature uniformity in the working volume.
- Capacity to handle the sample’s thermal load, especially if the product dissipates power.
- Configuration: a single-compartment chamber or a multi-zone chamber for thermal shock.
What the engineer sees in practice
The value shown on the display is the reading from a chamber sensor, not necessarily the sample temperature. That is why a thermocouple is placed on the part. A heavy sample or one that dissipates power can take a long time to reach the air temperature and lags behind the profile. An excessive load obstructs the airflow and worsens uniformity.
Where it falls short
A climatic chamber does not replace other tests: it does not apply vibration and does not generate salt fog or solar radiation. When the failure mechanism under study depends on several factors at once, combined chambers are needed, for example with integrated vibration. And humidity, at low or high temperatures, has physical limits that are shown on the operating chart of each piece of equipment.