What does radiometric calibration mean?

It means establishing, against reference sources of known temperature, the relationship between the signal the detector delivers and the temperature the camera must display, so that each pixel is a measurement and not just an image.

From image to measurement

A thermal camera can be imaging-only, showing contrasts without assigning reliable temperatures, or radiometric, where each pixel has an associated temperature. What sets them apart is radiometric calibration.

What it consists of

  1. A reference source, normally a blackbody, is set to several known temperatures within the camera’s interval.
  2. The detector signal is recorded at each point.
  3. The parameters of the function that converts signal into temperature are determined, including the effect of the optics and of the instrument’s internal temperature.
  4. The reading is verified against the reference and the deviation is documented.

Because the relationship between radiation and temperature is not linear, a calibration function is required, not a simple factor.

What it includes and what it does not

  • It includes the response of the detector, optics, and electronics.
  • It does not include the emissivity of the measured object, which the user adjusts in each case.
  • It does not include compensation for distance, humidity, and reflected temperature, which are measurement parameters.

Traceability

Confidence in the result depends on the reference itself being traceable to recognized temperature standards and on its uncertainty being known. A calibration without a reliable reference puts down a figure but gives no assurance.

How to interpret the result

  • A calibration certificate or report gives the instrument’s deviation at temperature points; it is valid for the stated conditions.
  • Calibration is a state at a point in time; wear, impacts, or a lens change can alter it.
  • An intermediate check with a simple reference helps detect drift between calibrations.

What it is not

Adjusting palette, level, or range is display only. Correcting emissivity in the menu is a measurement correction. Neither of the two is radiometric calibration.