Low-Noise Amplifier or Power Amplifier?

A low-noise amplifier is placed at the start of a receiver to add minimal noise to weak signals; a power amplifier is placed at the end of a transmitter to deliver the required output level with the necessary linearity and efficiency.

Two different problems

In a receiver, the signal arrives very weak and each stage adds noise. In a transmitter, the signal is already strong and what matters is bringing it to the output level without distorting it. Both are amplifiers, but they are optimized for opposite goals.

Low-noise amplifier (LNA)

Its main parameter is noise figure. The Friis formula shows that, in a chain, the noise of the first stage dominates: the noise of later stages is divided by the accumulated gain of those before them. That is why an LNA with sufficient gain at the input reduces the total noise figure of the receiver. Its input match is designed for minimum noise, which normally does not coincide with the match for maximum power transfer, and so its input return loss is a deliberate compromise.

Power amplifier (PA)

Here what matters is output power, efficiency and linearity. A PA operates near compression: as the input level increases, the gain stops being constant, and harmonics and intermodulation products appear. This is described by the compression point and the third-order intercept point. Efficiency determines how much power is converted into heat, so the heat sink and the biasing are part of the design.

Comparison

Aspect LNA PA
Location Receiver input, close to the antenna Transmitter output, before the antenna
Objective Minimum added noise Power and linearity
Input match For minimum noise For gain and power
Key measurement Noise figure and gain Compression, intermodulation, efficiency
Risk Saturating with strong signals Distortion and overheating

What is observed in measurements

An LNA is characterized with a calibrated noise source and an analyzer, or with the Y-factor method, because with weak signals its own noise is what is being measured. A PA is characterized with power sweeps and by analyzing the output spectrum, using attenuators and couplers so that the measuring equipment is not damaged.

Using one where the other belongs produces typical failures: a PA at the start of a receiver degrades sensitivity with its noise; an LNA in a transmitter compresses and distorts easily. Work with high radio-frequency power can cause burns and dangerous exposure, and must be carried out by qualified personnel following the applicable regulations.