Fundamentals
Volume resistivity characterizes how strongly the material opposes conduction through its interior. It is measured with a DC voltage between two electrodes and calculated from the resistance, the effective electrode area and the thickness. Because insulators have enormous resistances, the current is very small and an electrometer or picoammeter is required.
Why the guard is essential
The volume current flows between the electrodes, but so does a current along the surface of the sample. To measure only the former, the measuring electrode is surrounded by a guard ring at the same potential, which diverts the surface current to ground without passing through the instrument. Without a guard, the reading includes surface conduction and the apparent value comes out lower.
The electrification time
When a DC voltage is applied, the current is not constant: at first it contains the capacitive charging and the slow polarization of the material, and it decays over time toward a conduction value. That is why the procedure sets an electrification time before reading. Comparing values measured with different times is meaningless.
Factors that alter the result
- Humidity and temperature: resistivity falls as both rise, sometimes by several orders of magnitude.
- Applied voltage: in some materials the current is not proportional to the voltage.
- Electrode contact: poor contact with the sample adds error in materials with a rough surface.
- Noise and leakage in the setup: the wiring must be shielded and the support insulators of good quality.
How to interpret the result
Volume resistivity is expressed in ohm-meters or ohm-centimeters; always check the unit. A very high value indicates good insulation, but what matters in service is the value under real humidity and temperature conditions. Report the electrification time, the voltage and the ambient conditions along with the number.
When it does not apply
For semiconducting or dissipative materials, the measurement regime and the electrodes are different, and methods specific to those materials are used.