How is the shielding effectiveness of a material measured?

The signal transmitted through a sample of the material is compared with the signal transmitted without it, in the same setup, and the difference in decibels is the shielding effectiveness. The most widely used methods are the coaxial line with a sample holder and the chamber with a window in which the sample is placed.

Measurement principle

The method is always comparative. First, the transmission between a source and a receiver is measured with a reference (no material, or an open window). Then the sample is inserted and the measurement is repeated. The difference between the two, in decibels, is the shielding effectiveness of the material at each frequency. A network analyzer or a generator with a receiver is normally used.

Common setups

  • Coaxial sample holder (transmission line): the sample, in the form of a disk or ring, is placed inside a flared coaxial line. It is reproducible and useful for comparing materials, but it is limited to small samples and a restricted frequency range.
  • Chamber or box with window: one wall has an aperture over which the sample is fixed. A transmitting antenna is on one side and a receiving antenna on the other. It allows larger samples and materials such as fabrics, paints or sheets.
  • Two chambers coupled through a window: similar, with greater dynamic range thanks to the isolation between the chambers.

Precautions that determine validity

  1. Electrical contact between the sample and the holder must be continuous; a leak at the edge is measured as if it were a property of the material.
  2. The dynamic range of the system must comfortably exceed the expected effectiveness; if the sample attenuates more than the instrument can resolve, only a lower limit is obtained.
  3. The reference and load conditions must be identical except for the sample.
  4. It must be stated whether the electric field, magnetic field or plane wave is being measured, because the result depends on it.

How to interpret the result

The figure describes the material in a setup, not the enclosure in which it will be used. Seams, covers, ventilation and cables determine the final effectiveness, which is almost always lower than that of the material. For this reason, material measurement serves for selection and comparison, and the complete enclosure is evaluated separately.

A curve versus frequency with marked dips usually indicates resonances of the setup, not of the material. It is advisable to repeat the measurement with another sample from the same batch to find out what part of the value is real variability.