The challenge
Measuring the temperature of a surface with a contact sensor means that the sensor must reach the temperature of the surface and must not change it when touching it. Both conditions fail easily, because there is a thermal contact resistance between the sensor and the surface, and because the sensor, with its body and leads, draws heat away or adds heat.
What determines measurement quality
- Thermal contact. A rough, dirty or oxidized surface leaves insulating air gaps. A suitable thermal paste or adhesive reduces the contact resistance.
- Losses through the sensor. The sensor and its leads conduct heat to the surroundings, like a fin. If the sensor is bulky or the leads are thick, the reading falls between the surface temperature and the air temperature.
- Exposure to air. A sensor exposed to air currents or radiation cools or heats on its own.
- Thermal response. A sensor of greater mass takes longer to match the temperature and also cools the area where it rests.
Types of contact sensors
Fine-wire thermocouples, welded or bonded to the surface, probes with a flat or spring-loaded tip, and thin-film resistance sensors are used. Fine-wire thermocouples have little mass and respond quickly; handheld probes are practical but more susceptible to contact error.
Good practices
- Clean the surface and make sure it is flat in the contact area.
- Fix the sensor with constant, reproducible pressure, without deforming the material.
- Cover the sensor or insulate it from the surrounding air, so that its body is at the surface temperature.
- Run the lead along the surface for some distance before routing it away, so that it equilibrates with the surface.
- Wait for the reading to stabilize.
Common errors
- Resting the probe dry on a rough surface and taking the reading immediately.
- Using a large sensor on a small or thin part, which ends up cooled by the sensor itself.
- Ignoring the effect of air currents.
- Comparing with an infrared reading without considering that it depends on emissivity.
What the engineer sees
If the reading rises slowly without stabilizing, or varies with the pressure applied to the probe, contact is the problem. If the reading stays consistently below the temperature of a hot surface, heat is being lost through the sensor.
On very hot, energized or moving surfaces, contact measurement is performed by qualified personnel with the applicable protective measures.