General idea
Radiated emission is the electromagnetic field that equipment produces in space. It is measured at a controlled site, at a fixed distance, because the field varies with direction and distance. What matters is the maximum value in any direction and polarization.
Elements of the measurement
- Antenna: converts the field into voltage; its antenna factor relates the two. Different broadband antennas are used depending on the frequency range.
- Receiver or analyzer: measures with the established bandwidth and detectors.
- Turntable: rotates the equipment to find the direction of maximum emission.
- Antenna mast: varies the height, because the reflection from the ground plane adds to or subtracts from the direct wave.
Procedure
- Measure the ambient environment with the equipment off.
- With the equipment on, carry out a quick sweep with a peak detector in both polarizations.
- Identify the frequencies close to the limit.
- For each one, rotate the equipment and sweep the antenna height until the level is maximized.
- Take the final measurement with the required detector.
- Calculate the field by adding the antenna factor and the cable losses.
Measurement distance
Limits are defined for a specific distance. If the measurement is made at another distance, it is sometimes permitted to convert the limit by assuming that the field decreases inversely with distance, but this conversion is an approximation valid under far-field conditions and does not apply across the whole range; it must be checked in the current standard. At shorter distances the measured level increases, but so does the influence of the near field.
How to interpret the result
An emission that rises and falls as the antenna height changes is normal; what matters is the maximum. A signal present with the equipment off is ambient and must be excluded. If two cable orientations change the result, the arrangement must be fixed and documented.