How is immunity to radiated fields measured?

A uniform, modulated electromagnetic field is generated with an antenna and an amplifier and swept in frequency over the equipment, while observing whether its operation stays within the defined performance criterion. The field strength is calibrated beforehand with a probe in the test area.

What is evaluated

Radiated immunity testing checks whether equipment keeps working when exposed to radio-frequency fields from nearby transmitters: portable radios, cellular telephony, broadcasting, industrial equipment. It does not measure a value of the equipment but its behavior in a known field.

Test chain

  • Signal generator with amplitude modulation.
  • Broadband power amplifier.
  • Transmitting antenna inside an anechoic or semi-anechoic chamber with absorbers.
  • Isotropic field probe to verify the field strength.
  • Equipment monitoring, by camera or through the functional output.

Field calibration

Before testing, a uniform field plane is calibrated at the location where the face of the equipment will be. Calibration is done without the equipment and verifies that the field strength is sufficiently uniform over a grid of points. Then, during the test, the system uses the calibrated power for each frequency.

Procedure

  1. Place the equipment at the defined distance and height, with its representative cabling.
  2. Sweep the frequency range in small steps, with a dwell time long enough for the equipment to respond.
  3. Repeat with the antenna in horizontal and vertical polarization.
  4. Repeat with each face of the equipment oriented toward the antenna.
  5. Observe and record any deviation from the expected behavior.

What the engineer sees

Symptoms include erroneous readings on analog sensors, noise on audio, resets, communication errors or oscillating outputs. Modulation matters because non-linear circuits demodulate the carrier and the equipment reacts to the envelope.

How to interpret the result

It is classified according to the performance criterion of the product standard. A failure localized at one frequency and orientation indicates the entry path: cable, enclosure aperture or sensitive trace. With that information, the designer can correct the exact point.