The problem with what rarely happens
A rare event is difficult because you cannot watch the screen all the time and because, between captures, the instrument spends time processing during which it sees nothing. The strategy has two parts: define the fault precisely so that the instrument can recognize it, and make sure the capture preserves the context.
How to define the event
- Runt, pulse width or timeout trigger: capture pulses of the wrong size or the absence of an expected edge.
- Level or window trigger: detects when the voltage leaves an allowed range.
- Protocol trigger: a frame error, or a specific address or data value.
- Mask test: compares each waveform with a template and stops on or flags those that fall outside it.
How to preserve the evidence
- Use pretrigger to see what happened before the event, which is usually where the cause lies.
- Use deep memory or segmented memory, which stores many short segments and reduces the dead time between triggers.
- Set the instrument to stop on trigger, or to save the capture automatically.
- Turn on persistence to accumulate images and see whether there are occasional different shapes.
Correlation
Recording the faulty signal at the same time as other candidates, such as the power supply, the clock or the input, lets you see what changed just before. If the fault depends on temperature, vibration or a software state, note those conditions alongside the record.
What to check first
Before waiting for hours, check that the trigger works by forcing a known fault, for example by touching a wire or applying a test pulse. A trigger that does not detect the simulated fault will not detect the real one either. If the system fails when you connect the probe, the problem may be one of contact or loading, and the probe itself may be modifying it.