How is the moisture absorption of an insulator evaluated?

The moisture absorption of an insulator is evaluated by comparing the mass of a dry specimen with its mass after exposure to water or to a defined humid atmosphere, and it is expressed as the mass increase relative to the dry mass.

Principle of the method

The gravimetric method is the most direct. The specimen is dried to a stable mass, that mass is recorded, the specimen is exposed to the chosen condition and it is reweighed at intervals. The mass increase divided by the dry mass, normally as a percentage, is the absorption. The method is simple, but it requires a suitable balance and careful handling.

Two types of exposure

  • Water immersion: reports the maximum absorption under liquid contact. It is useful for comparing materials and detecting sensitivity to water.
  • Controlled-humidity atmosphere: comes closer to the real operation of equipment in a humid environment and can give equilibrium values different from those of immersion.

Diffusion sets the time

Water enters by diffusion, so the time to reach equilibrium grows with the square of the thickness. A thin specimen stabilizes quickly, and a thick one can take a long time. The curve of mass against the square root of time is usually linear at the start if the process is simple diffusion, and that slope makes it possible to estimate a diffusion coefficient. If the curve does not level off, there may be loss of soluble components, hydrolysis or degradation.

Measurement precautions

  1. Dry the specimen and keep it in a desiccator while it cools.
  2. Remove surface water before each weighing, always in the same way.
  3. Weigh quickly to avoid evaporation.
  4. Bear in mind that edges and faces contribute different absorption areas.

From mass to electrical properties

Absorption is an indirect indicator. Water in the material raises the permittivity, increases dielectric losses and reduces resistivity and surface dielectric strength. A small percentage of water can produce notable electrical changes, so the relationship is determined by measuring the electrical property on specimens with different water contents, not by assuming it.

What to look for in the result

Compare values from materials with identical geometry and exposure. Check whether the process is reversible on drying: if the mass does not return to its initial value, there were chemical changes or loss of additives.