The challenge
An event that does not repeat cannot be averaged or reconstructed; the instrument must see it at the moment it occurs. You have to decide in advance which trigger identifies it, how much detail is needed, and how much context before and after.
Step by step
- Estimate the event: approximate amplitude, duration and what precedes it. This defines the vertical scale, the sampling speed and the time window.
- Adjust the vertical scale so that the normal and abnormal signals both fit without saturating. If the event is a peak larger than the normal signal, too sensitive a setting clips it.
- Set the sampling and the memory: the rate must be sufficient for the fastest expected edge, and the memory must cover the complete window at that rate.
- Choose the trigger: an edge with a level just above the noise, or something more specific, such as pulse width, short pulse or pattern, if the event has a distinguishable signature.
- Position the trigger: move the reference to show more time before the event, which is what helps uncover the cause.
- Arm and wait: in single mode the instrument captures once and stops; in normal mode it updates on each valid event.
Useful tools
- Segmented memory, which stores many consecutive events with their time stamps, without capturing the quiet intervals.
- Mask or limit triggering, which reacts when the signal leaves an allowed region.
- A second input that records another related signal, such as a power supply or a control line, to find the cause.
How to interpret the result
If it does not trigger, the level may be wrong, the source may be a different one, or the event may not occur at the measurement point. A convenient test is to force a sample event with a generator before leaving the equipment waiting, to validate the entire setup.
When the signal belongs to energized or power equipment, connecting the probes is a job for qualified personnel, in accordance with the applicable regulations.