Definition
Group delay is the negative of the derivative of the transmission phase with respect to angular frequency. It describes the time the envelope of a signal with frequencies close to the one considered takes to pass through the device. If it were constant across a band, all components would be delayed equally and the signal would pass without phase distortion.
How it is measured with a network analyzer
- Calibrate the transmission measurement with the reference plane at the filter connectors.
- Measure S21 over a frequency range that covers the band of interest with enough points.
- Request the group delay trace from the instrument, which calculates the phase difference between neighboring points divided by the frequency difference.
The aperture
The calculation uses a frequency interval called the aperture. With a very small aperture, phase noise dominates and the trace is noisy. With a very large one, rapid variations are averaged and details are lost. The aperture should be chosen so that the phase changes moderately between points and the details that matter are visible. The instrument must be able to unwrap the phase without 360 degree jumps, which requires a sufficiently fine frequency step.
What is seen in a filter
- At the center of the passband the delay is relatively flat.
- Near the edges it rises sharply, and more so the more selective the filter is.
- Outside the band the delay has no practical meaning, because the signal is heavily attenuated and the phase is noisy.
Why it matters
In digital communication systems, non-uniform delay distorts pulses and causes intersymbol interference. In radar, it deforms compressed pulses. For this reason it is preferable to know the delay variation across the band in use.
How to interpret the result
What matters is less the absolute value than its variation within the channel. Cables and accessories add a constant delay that can be subtracted with an electrical length compensation without affecting the shape.