What is actually measured
Insulation resistance is the ratio between an applied DC voltage and the current passing through the insulator, adding the current through the volume of the material and the current flowing across its surface. Humidity increases both: absorbed water raises the volume conductivity, and the surface film, with ionic contaminants, opens leakage paths.
General procedure
A baseline is measured before exposing the sample, under controlled environmental conditions. The measurement is then repeated during exposure, if the setup allows, and after it, with a defined recovery period. Comparing the three stages separates a reversible effect, due to surface moisture, from a permanent one, which indicates material damage.
What must be kept constant
- The test voltage and its polarity.
- The electrification time before reading, because the current decreases over time as dielectric absorption dissipates and the reading depends on the instant chosen.
- The sample temperature, since insulation resistance falls as temperature rises.
- The connection points and the cleanliness of cables and sample holders.
Sources of error
Leakage in cables, probes and supports can dominate the reading of a high resistance. An electrometer or a high-resistance meter with a guard is used, with clean, dry wiring. Measuring immediately after removing the sample from the chamber is also misleading: condensation on the outside can cause leakage that does not represent the material.
How to interpret the result
The trend matters more than a single value. A large drop during exposure that recovers on drying points to surface moisture or reversible absorption. A drop that does not recover suggests chemical degradation, delamination or contamination. The acceptance criterion comes from the product specification and must be set before the test.
When it does not apply
If the design requires knowing the breakdown voltage or the resistance to leakage currents under a high electric field, a different test is appropriate. The insulation test voltage can be dangerous; measurements at high voltages must be performed by qualified personnel following the applicable safety regulations.