Operating principle
Compressed air, humidified and sometimes heated, passes through a nozzle that atomizes a salt solution. The fog spreads through a closed enclosure whose temperature is kept under control. The droplets settle on the samples and form a continuous electrolyte film that promotes corrosion reactions. Excess condensate is drained.
Main components
- Solution reservoir and water preparation system.
- Atomizer and conditioned air supply.
- Enclosure with a water-sealed or gasketed lid.
- Heating of the enclosure and of the air humidification.
- Collectors for measuring fog deposition.
What is controlled
The concentration of the solution, its acidity, the temperature and the amount of fog deposited per unit of area and time. If these parameters drift, the aggressiveness changes and the results are not comparable. They are therefore verified with collectors and reference samples, not just with the chamber sensors.
Variants
The continuous version maintains uninterrupted fog. The cyclic versions alternate fog, drying and humidity, and come closer to what happens in service, where coatings get wet and dry out. The drying cycles matter because atmospheric corrosion advances markedly during the wetting and drying stages.
How to interpret the result
The test is comparative and serves for control. It makes it possible to compare coatings, detect application defects or verify that a process remains constant. There is no fixed factor that converts hours of salt spray into years of service, because real corrosion depends on the climate, the contaminants and the geometry of each part. That equivalence is best avoided.
Practical care
Samples must not touch each other or drip onto one another, and their position relative to the fog must be reproducible. The chamber requires regular cleaning, because salts clog nozzles and alter the deposition. The chemicals and the solution are handled with suitable protective equipment.