Peak, quasi-peak or average detector?

The peak detector records the maximum value and is the fastest and most conservative; the average detector reflects the mean value of the envelope and is useful for continuous narrowband signals; the quasi-peak detector weights amplitude according to repetition rate. The following always holds: peak is greater than or equal to quasi-peak, and quasi-peak is greater than or equal to average.

What each detector does

  • Peak: follows the signal envelope and holds the maximum value during the measurement time. It does not depend on how many times the pulse occurs.
  • Quasi-peak: a detector with charge and discharge time constants defined by the standard. An isolated, infrequent pulse discharges between events and produces a low reading; repetitive pulses keep the detector charged and the reading rises. It mimics the annoyance effect of interference on radio reception.
  • Average: responds to the mean value of the envelope. It tends to ignore impulsive noise and to highlight continuous carriers.

How they relate

For a given signal, the peak reading is greater than or equal to the quasi-peak reading, and the quasi-peak reading is greater than or equal to the average reading. For an unmodulated continuous carrier, all three are practically equal. For low-repetition impulsive noise, peak is much higher than average.

How they are combined in practice

  1. A fast sweep is made with the peak detector, because it is the fastest and most conservative.
  2. If the peak is below both the quasi-peak and the average limit, the equipment complies and there is no need to continue.
  3. If the peak exceeds either limit, only those frequencies are remeasured, with the detector the limit requires.
  4. Each result is compared with the limit for the same detector.

How to interpret the results

Comparing a peak reading with an average limit can suggest a noncompliance that does not exist; it is the most pessimistic reading. Conversely, comparing an average reading with a quasi-peak limit would give an overly optimistic conclusion. Each limit is defined for a specific detector.

Equipment with continuous, narrowband noise, such as a clock or a switching power supply, shows almost the same reading on all three, so the chosen detector changes the result very little. Equipment with a brushed motor generates irregular bursts, and the difference between detectors can be large.